🔸Mass Flowmeter

🔹 Gas Mass Flow Meter

• Directly measures mass flow with high measurement accuracy.
• Measures a wide range of fluids, including various liquids with high viscosity, slurries containing solids, liquids with trace gases, and medium-to-high pressure gases with sufficient density.
• The measuring tube exhibits minimal vibration amplitude and can be considered a non-moving part, with no obstructions or moving components within the measurement pipeline.
• Insensitive to counter-current velocity distribution, eliminating requirements for upstream/downstream straight pipe sections.
• Minimal impact of fluid density variations on measurement values.
• Capable of multi-parameter measurement, such as simultaneous density measurement, enabling derived concentration measurements of solutes in solutions.

I. Introduction

In modern industrial production and scientific research, the precise measurement of fluid mass flow is of vital importance. The Coriolis Mass Flowmeter, as a high-precision flow measurement instrument, is widely used in multiple industries such as petroleum, chemical engineering, food, and pharmaceuticals. It operates based on the Coriolis force principle and can directly measure the mass flow of fluids, unaffected by the fluid's temperature, pressure, viscosity, and other physical properties.

 

II. Principles

1. Generation of Coriolis Force

- The Coriolis force is an inertial force that occurs in a rotating reference frame. When a particle moves in a straight line relative to a rotating reference frame, it experiences a force perpendicular to both its direction of motion and the axis of rotation. This is the Coriolis force.

- The magnitude of the Coriolis force can be expressed by the formula F = 2mv × ω, where m is the mass of the particle, v is the velocity of the particle relative to the rotating reference frame, and ω is the angular velocity of the rotating reference frame.

2. Measurement Principle of Coriolis Mass Flowmeter

- The Coriolis Mass Flowmeter consists of one or more vibrating tubes. When fluid flows through the vibrating tubes, the tubes are driven to vibrate at a fixed frequency.

- According to the Coriolis force principle, when fluid flows inside the vibrating tubes, it generates a Coriolis force, causing the tubes to twist.

- The mass flow of the fluid is directly proportional to the degree of twist of the vibrating tubes. By measuring the degree of twist of the tubes (typically through sensors installed on the tubes to measure phase difference or frequency change), the mass flow of the fluid can be calculated.

 

III. Structure

1. Sensors

- The sensor is the core component of the Coriolis Mass Flowmeter, including the vibrating tube, driver, and detection element.

- The vibrating tube usually adopts U-shaped, Ω-shaped, or straight structures, and its material is selected based on the properties of the fluid being measured, such as stainless steel or Hastelloy.

- The driver is used to make the vibrating tube vibrate, with common driving methods being electromagnetic and piezoelectric.

- The detection element measures the vibration characteristics of the vibrating tube, such as strain gauges or capacitive sensors, obtaining flow information by detecting the displacement, phase difference, or frequency change of the vibrating tube.

2. Transmitter

- The transmitter is responsible for processing, amplifying, and converting the signals collected by the sensor, and outputting them as standard electrical signals (such as 4 - 20mA or digital signals).

- The transmitter typically contains signal conditioning circuits, microprocessors, memory, and communication interfaces, capable of performing flow measurement, data processing, fault diagnosis, and communication functions.

3. Housing and Connection Components

- The housing protects the sensor and transmitter, usually made of metal or engineering plastics, with good sealing and corrosion resistance.

- Connection components include pipe connection flanges or threaded interfaces, used to connect the flowmeter to the fluid pipeline, ensuring smooth flow of the fluid into and out of the flowmeter.

 

IV. Applications

1. Petroleum and Natural Gas Industry

- In the petroleum refining process, Coriolis Mass Flowmeters are used to precisely measure the mass flow of crude oil, refined oil products (such as gasoline, diesel, lubricating oil, etc.), ensuring accurate ingredient mixing and product quality control in the production process.

- In natural gas transportation and metering, it can measure the mass flow of natural gas, unaffected by changes in natural gas pressure, temperature, and composition, ensuring fair and accurate trade settlement. In chemical production, for the measurement of the mass flow of various chemical raw materials (such as acids, alkalis, salt solutions, organic solvents, etc.) and products (such as plastics, rubber, fertilizers, etc.), Coriolis mass flowmeters can provide highly accurate measurement results, which helps optimize production processes and improve product quality.

3. Food and Beverage Industry

- In the food processing industry, on production lines for liquid foods such as milk, juice, and beer, Coriolis mass flowmeters are used to precisely measure the addition of raw materials and the packaging volume of products, ensuring the consistency of product quality and compliance with hygiene standards.

4. Pharmaceutical Industry

- The pharmaceutical industry has extremely high requirements for the accuracy and hygiene of flow measurement. Coriolis mass flowmeters can accurately measure the mass flow of pharmaceutical water, drug solutions, and various additives, ensuring precise control of the drug production process and the stability of drug quality.

V. Advantages

1. High Precision Measurement

- Coriolis mass flowmeters can directly measure the mass flow of fluids with high accuracy, generally reaching ±0.1% - ±0.5%, and even higher in some high-precision applications.

2. Unaffected by Fluid Physical Properties

- Their measurement results are not affected by changes in the fluid's temperature, pressure, viscosity, density, and other physical properties, providing stable and reliable measurement data.

3. Multi-parameter Measurement Capability

- In addition to measuring mass flow, they can also simultaneously measure the density and temperature of the fluid, providing more parameter information for process control and quality monitoring.

4. No Moving Parts

- There are no moving parts inside the flowmeter, reducing mechanical wear and failure probability, and improving the reliability and service life of the equipment.

VI. Limitations

1. High Cost

- Due to its high precision and complex technical structure, the manufacturing cost and purchase price of Coriolis mass flowmeters are relatively high, which may be a limitation in some cost-sensitive applications.

2. Large Pressure Loss

- When the fluid flows through the vibrating tube, a certain pressure loss occurs. For systems with strict requirements for pressure loss, its applicability needs to be comprehensively considered.

3. Strict Installation Requirements

- During installation, it is necessary to ensure that the flowmeter's pipeline is completely filled with fluid and avoid the influence of pipeline vibration and stress on the flowmeter, otherwise it will affect the measurement accuracy.

VII. Conclusion

Coriolis mass flowmeters have been widely applied in many industries due to their high precision, direct measurement of mass flow, and insensitivity to fluid physical properties. Although there are limitations such as high cost, large pressure loss, and strict installation requirements, these issues are gradually being improved with the continuous advancement of technology. In applications where high accuracy and reliability of flow measurement are required, Coriolis mass flowmeters are undoubtedly an ideal choice.


Product Parameters


Medium under test: Liquid, gas 

Measuring diameter: 3mm to 200mm 

Measurement items: Mass flow rate, density, temperature, volumetric flow rate, % concentration 

Density measurement range: 0.2 g/cm³ to 2.5 g/cm³ 

Medium temperature: -200℃ to +350℃ 

Flow measurement error: 0.15%R + Cz for liquid, 0.2%R + Cz for liquid, 0.5%R + Cz for gas, 1.0%R + Cz for gas. 

Density measurement error: 0.002 g/cm³ 

Temperature measurement error: ≤ 1℃ 

Signal form: 4~20mA, 0~10KHZ, RS485, Hart 

Power supply: 24VDC 

Material of the measuring tube: 316L. Also, it can be provided as per the user's requirements through negotiation. 

Explosion-proof grade: ExdibIICT6, ExdIICT4 

Enclosure protection grade: IP65, IP67 

Flange standards: National standard GB/T series standards, and can also be processed according to user requirements.


a. General Installation Requirements for Flow Sensors

Due to the differences in the shape and structural design of the measuring tubes, the installation requirements vary greatly. Therefore, it is necessary to follow the installation methods and precautions specified by our factory. For example, some models of flow sensors can be directly connected to the pipeline, while others require the installation of support frames or bases. To isolate the influence of pipeline vibration on the instrument, sometimes a flexible pipe is used to connect the sensor and the pipeline, and a rigid straight pipe with a support is used between the flexible pipe and the sensor. Before purchasing, please request the installation and operation manual from our company for reference, comparison, and selection.

When designing the installation, it is necessary to ensure a long service life. To remove solid particles and entrained gases that may cause premature wear and measurement errors, install a filter or gas separator and other protective devices upstream of the sensor according to the fluid and pipeline conditions. If you want to calibrate the instrument on-site, consider the connection port and valve for drainage, as well as the corresponding space.

 

 

b. Installation Posture and Position of Flow Sensors

Residual solid particles, fouling, and retained gases in the measuring tube of the flow sensor will affect the measurement accuracy. Generally, it is ideal to install the sensor in a vertical pipe with upward flow. However, for non-straight measuring tube mass flowmeters, whether to install them in a vertical or horizontal pipe depends on the vibration conditions and application requirements of the pipeline. The installation position must ensure that the measuring tube is filled with liquid. For example, if the fluid flows through the mass flowmeter and is directly discharged into a container without back pressure on a horizontal pipe, the measuring tube is often not filled, causing the output signal to fluctuate violently.

 

 

c. Installation of Shut-off Valves and Control Valves

To ensure no flow during zero adjustment, shut-off valves should be installed upstream and downstream of the mass flowmeter and must be leak-free. The control valve should be installed downstream of the mass flowmeter to maintain as high a static pressure as possible in the mass flowmeter to prevent cavitation and flashing.

 

 

d. Pulsation and Vibration

To prevent mechanical vibrations in the process and from the outside from affecting the mass flowmeter, inquire with our factory about the resonance frequency range of the provided mass flowmeter to determine if the vibration frequency at the site is close to the resonance frequency of the mass flowmeter. You can also provide the vibration conditions at the site to our factory to consult whether the following measures are necessary: 1) installing a pulsation damper, 2) installing a vibration damper or flexible connection pipe, 3) using special clamping and fixing equipment for the flow sensor, etc.

 

 

e. Preventing Mutual Influence Between Mass Flowmeters

When two mass flowmeters of the same model are installed in series or multiple mass flowmeters are installed in parallel (or in parallel) close to each other, especially on the same support frame, the vibration of the measuring tubes can cause mutual influence between the mass flowmeters, generating interference and causing abnormal vibration. In severe cases, the instrument may not work. Preventive measures should be taken during installation, such as: 1) requesting our factory to adjust the resonance frequency values of the adjacent instruments, 2) increasing the distance between the flow sensors, not installing them on the same frame, and setting up independent support frames, 3) installing the flow sensors in different directions, 4) installing vibration isolation materials between the flow sensors, etc. 

 

f. Pipeline Stress and Distortion

When tightening the bolts to connect the flange of the mass flowmeter with the pipeline flange, do it evenly to avoid applying stress to the mass flowmeter (e.g., due to the two flange planes of the pipeline not being parallel). If a short pipe of the same length as the mass flowmeter is pre-installed during pipeline layout, it can prevent stress caused by improper pipeline layout. During operation, due to changes in process pressure and temperature, the mass flowmeter may be subjected to axial force or bending/twisting force from the pipeline, affecting measurement performance. Appropriate support frames should be provided.

 

 

g. Use of Strongly Abrasive Slurries

When measuring strongly abrasive slurries, it is best to use a straight-tube single-tube type and keep the measuring tube in a vertical position to avoid uneven wear of the pipe wall and shorten the service life. However, thinning of the pipe wall will reduce the rigidity of the measuring tube and change the flow measurement value. Therefore, during operation in such environments, regular inspections should be conducted to confirm the service cycle. Scaling or drift deposits on the inner wall structure of the measuring tube can also affect measurement accuracy, so regular cleaning is necessary.


h. Zero drift and zero adjustment

Zero drift comes from the flow sensor part. The main reasons are: 1) Asymmetry and attenuation of mechanical vibration; 2) Changes in fluid density and viscosity. The factors affecting the former are: a) The influence of fixed stress at the pipe end; b) Changes in the stiffness of the vibrating tube; c) Inconsistency in the resonant frequency of the dual tubes; d) Internal attenuation of the pipe wall material. The reason for the latter affecting the zero position is structural imbalance. Therefore, even if the resonant frequencies of the dual tubes are adjusted to be consistent when the tube is empty, zero drift may occur when the tube is filled with liquid. Similarly, the vibration attenuation caused by viscosity is related to frequency and may also cause zero drift during flow.


Finally, zero adjustment must be carried out on-site after installation. The flow sensor should be purged of gas and filled with the fluid to be measured. Then, the upstream and downstream valves of the sensor should be closed, and zero adjustment should be performed under conditions close to the working temperature. Re-adjustment is required when there are changes in installation or significant temperature variations.

 


Installation

- Location Selection: Install away from large motors, variable frequency drives, and other sources of interference to prevent electromagnetic interference from affecting measurement accuracy. Avoid locations with significant temperature fluctuations or strong mechanical vibrations.

- Pipeline Layout: Ensure sufficient straight pipe sections upstream and downstream to stabilize fluid flow. Install according to flow direction indicators, securing the unit firmly to prevent stress concentration damage.

Operation

- Fluid Compatibility: Verify fluid properties (e.g., density, viscosity, corrosiveness) fall within the flowmeter's specifications. Avoid measuring fluids containing excessive bubbles or solid particles unless specially treated.

- Operational Guidelines: During startup and shutdown, adjust flow rate gradually to prevent water hammer impacts. Monitor fluid temperature and pressure during operation, ensuring they remain within specified limits.

Maintenance

- Regular Cleaning: Periodically clean the flow meter to remove internal impurities and deposits, preventing scaling that could affect measurement accuracy.

- Calibration and Inspection: Calibrate according to specified intervals to ensure measurement accuracy. During routine inspections, check the instrument's appearance and verify the integrity of connection lines.